Tel +(618) 9361 2228 Fax +(618) 9361 2216 | |
|
| |
|
|
|
|
TGA1000 Thermogravimetric Analyser

The TGA1000 Robotic Thermogravimetric Analyser system is a fully integrated system comprising of four furnaces, a loading station and a robotic manipulator. Thermal Analysis Testing is performed on samples which are automatically loaded into a carousel before being transported to the various furnaces by means of a robotic manipulator.
 |
SAMPLE FURNACES The TGA1000 Robotic Thermogravimetric Analyser is primarily used for thermal analysis testing on meterials such as Iron Ore, Coal and Organic materials. The furnaces are user configurable to temperatures ranging from ambient to 1000oC and are supplied with an integrated gas purging and extraction systems. |
|
SAMPLE LOADING & DISPENSING SYSTEM The TGA1000 Thermogravimetric Analyser receives a vial of sample from an automated sample prepartion system. The TGA1000 then takes a sample from the vial and place it into a cleaned crucible. Once the sample has been weighed, the entire carousel is then transported by a robotic manipulator to the first furnace. The TGA will measure the Loss on Ignition (also known as thermal analysis testing) at this first temperature.
Once the Loss on Igintion (LOI) is completed in the first furnace the carousel is then moved to the second furnace and LOI is re-measured at a higher temperature. The cycle is repeated for furnace three and four.
|
 |
 |
Cooling Stations The TGA1000 Robotic Thermogravimetric Analyser has four cooling stations where the crucibles and carousels are air cooled after thermal analysis in furnace four. The carousel is placed in the cooling cabinet for 15 minutes to cool. Thereafter the carousel is removed and sent to the loading station where the entire thermogravimetric cycle begins again. | TGA Control The TGA1000 Robotic Thermogravimetric Analyser system is controlled using stand alone TGA1000 software installed on a Windows XP Pro PC.
In-Feed Conveyors Two in-feed sample conveyors are included in our supply to transport the sample in cups to the TGA1000. These conveyors are fully automatic and are integrated within the robot control System.
Sample through put The four furnaces TGA1000 Robotic Thermogravimetric Analyser system can provide 40 four stage LOI results approximately every 30-35 minutes. This calculates to around 1,900 samples in a 24 hour working day.
Specifications of the TGA1000 Thermograimetric Analyser
| Power Requirements: |
|
| AC Supply |
3 Phase, 415 VAC 50Hz configurable on request |
| AC Current |
60 Amps per Phase |
| AC Max Power Consumption |
(14.4kW) per Phase or 43.2kW total |
| Furnaces: |
|
| Typical temperature range |
Furnace 1 105oC |
| Typical temperature range |
Furnace 2 370oC |
| Typical temperature range |
Furance 3 650oC |
| Typical temperature range |
Furance 4 1050oC |
|
|
| Gases: |
|
| Air Pressure |
100 PSI Dry, clean instrument air |
| Nitrogen Supply |
100 PSI, 10 Liters per minute |
| Oxygen Supply |
100 PSI, 10 Liters per minute |
|
|
| Footprint: |
|
| Width |
2310mm |
| Length |
5280mm |
| Height |
2025mm | |
|
|
|
|
|
|
|
|
|
|
|
 |
|
 |
|